Brain Matters Imaging & Treatment Centers®

NEUROSPECT QUAD Gamma Camera


Features:

  • Configured specially for brain imaging
  • Four Fixed 90° ring small FOV ultra high resolution detectors
  • Four independent energy channels
  • Excellent resolution / short time imaging
  • 2.5mm low attenuation SPECT pallet
  • High density FPGA technology
  • Ultra fast, 40 MHz digitizing
  • 4096 pixel image digitizing
  • Real-time cross correction for each energy level and different collimators
  • SW-based pile-up correction for high count rate
  • Pre-programmed robotic gantry motions for precise positioning
  • Ring Gantry Design
  • Fixed position SPECT gantry, high stability
  • Low energy high resolution (LEHR) collimator
  • 90° and 180° SPECT movements
  • Motorized vertical movement imaging table (250mm)
  • Specially design comfortable head support
  • Easy to change collimators
Options:
  • Additional collimators
  • Collimator cart
  • SPECT table accessories
Parameter
Central Field of View (CFOV)
Useful Field of View (UFOV)
Intrinsic spatial resolution
FWHM d 2.72mm d 2.8mm
FWTM d 5.44mm d 5.6mm
     
Intrinsic energy resolution
FWHM d 9.7 % d 9.7 %
     
Intrinsic flood field uniformity
Integral ± 2.4 % ± 2.9 %
Differential ± 1.9 % ± 2.4 %
     
Intrinsic spatial linearity
Absolute d 0.38 mm d 0.40 mm
Differential d 0.18 mm d 0.20 mm
     
Intrinsic high count rate performance
20 % count rate loss 160 kcps 160 kcps
maximum count rate 200 kcps 200 kcps
     
Intrinsic spatial resolution at 75 kcps
FWHM d 2.5 mm d 2.54 mm
FWTM d 4.97 mm d 4.97 mm
     
Intrinsic flood field uniformity at 75 kcps
integral ± 2.6 % ± 3.8 %
differential ± 1.9 % ± 2.2 %
     
Multiple window spatial resolution
  d 1.5 mm d 1.5 mm
     
Point source sensitivity
  ± 0.5 % ± 0.5 %
     
System spatial resolution w/ scatter (LEAP @ 10cm)
FWHM d 8.7 mm d 8.7 mm
FWTM d 15.5 mm d 15.5 mm
     
System spatial resolution w/o scatter (LEAP @ 10cm)
FWHM N/A N/A
FWTM N/A N/A
     
System sensitivity
LEAP Collimator 360 cpm/(micro)Ci 360 cpm/(micro)Ci
 
Performance Specifications - Reconstructed System Spatial Resolution
 
Radial
Tangential
Center
 
(Non RM2) SPECT
  d 7.5 mm d 4.0 mm d 7.5 mm  
Resolution Matched Matrix (RM2) SPECT
  d 5.6 mm d 3.2 mm d 5.6 mm  
 
33 (5.08cm diameter) high performance photo-multiplier tubes (PMTs)
Detector thickness 9.50mm / FOV 215 x 230mm